Calibration of Spectrophotometric Standards from
the Ground
Jonathan Burkle, Nick Mostek, Stuart Mufson
Indiana University
During this summer, REU student Jon Burkle (Wheaton College), working
in collaboration with Stuart Mufson and Nick Mostek, investigated whether
SNAP fundamental spectrophotometric standards could be calibrated from the
ground, relative to a light source, in narrow spectral bandpasses free of
water vapor and other sources of opacity. As a reminder, the SNAP fundamental
spectrophotometric calibrators, at least in the current calibration plan,
have not been measured relative to a calibrated light source in the near
infrared, but are assumed to be calibrated sufficiently accurately by stellar
atmospheres models. Our long range goal is to test whether these stellar
atmospheres models are consistent with the stellar fluxes in select narrow
band filters calibrated relative to a light source. If they are consistent,
then the SNAP collaboration can have confidence in the calibration of the
supernova data, which ultimately derives from the calibration of the fundamental
standards. If the fluxes are not consistent, then the SNAP collaboration
will need to invest more heavily in the near infrared calibration of the
fundamental photometric standards. The reason for the R&D described
here is to test whether such an experiment is even feasible.
We investigated the three bandpasses shown below. These bandpasses
all had commercially available interference filters that were purchased from
Oriel.
- 1050 nm, 10 nm bandpass: As seen below, this bandpass is virtually
free of opacity
- 1200 nm, 10 nm bandpass: This filter has approximately 1% attenuation
in its bandpass
- 1400 nm, 80 nm bandpass: This filter is heavily obscured by
water vapor
The R&D effort this summer had two components.
I. 1/R**2 Tests
We first measured a calibrated light source under constant temperature
and humidity conditions in the lab at both 7.8m and 8.5m ('near') to test
the accuracy at which we could measure 1/R**2. Our measuring apparatus
in the near infrared consisted of a single channel photometer equipped with
a Hamamatsu 3mm, thermoelectrically cooled InGaAs detector. For the
R-band measurements, a Hamamatsu Si-photodiode replaced the InGaAs detector.
The photometer was mounted on an F10 Meade 8-inch reflecting telescope.
The DAQ used Vernier LoggerPro hardware and software. The relative
humidity was constantly monitored with a Vernier humidity meter that was read
out simultaneously. The light source was a calibrated 45W QTH Oriel
light source powered by a regulated power supply. Laboratory measurements
by Nick Mostek using a calibrated Si photodiode and a calibrated InGaAs detector
were used to verify the lamp calibration. (See Nick's talks at the hardware
meetings for more information.)
In Fig.1 we show a scatter diagram of the laboratory 1050 nm bandpass
measurements (first 41 trials). What is shown is the fractional
error in the measurements (not % error):
[measured(flux at 8.5m/flux at 7.8m)**2 - expected (flux at 8.5m/flux
at 7.8m)**2]/expected(flux at 8.5m/flux at 7.8m)**2
Figure 1. Scatter diagram of the fractional error in the 1/R**2
'near' measurements at 150 nm
A histogram of the laboratory measurements (first 41 trials of the scatter
diagram) is shown in Fig.2(a). (the unreadable abscissa is now % error)
The distribution is reasonably Gaussian. We repeated the measurements
at 7.8m and 8.5m on the Swain Hall roof in conditions of > 95% relative
humidity. These measurements are shown in Fig.1 for trials 42 and
above. Although the measurements are systematically lower than the
lab measurements (i.e., less light than expected), the systematics in the
measurements most likely make the difference insignificant. The Swain
Hall roof is made up of tiles that move whenever people walk over them which
can cause the flux ratios to vary by 1-2% as the observer moves the light
source back and forth. In addition, the distribution of the measurements
do not differ significantly from a Gaussian.
Figure 2. (a) Laboratory 'near' measurements of 1/R**2 in the
lab at 1050 nm. (b) All 'near' measurements of 1/R**2 at 1050 nm (lab
+ roof)
In Table 1 we summarize the results of the 'near' measurements in
the lab through the 1050nm filter. In Table 2 we give the absolute flux
normalization at 7.8m in the 1050nm filter.
In Fig.3(a) we show the histogram of the measurements at 1400 nm
in the lab and in Fig.3(b) the measurements in the lab and on the roof of
Swain Hall. In this case the effect of water vapor on the ratios is
clearly apparent and statistically significant. These results are summarized
in Table 1. In Table 2, we give the absolute flux normalization at 7.8m
in the 1400nm filter.
Figure 3. (a) Laboratory 'near' measurements of 1/R**2 in the
lab at 1400 nm. (b) All 'near' measurements of 1/R**2 at 1400 nm (lab
+ roof)
The measurements at 1200nm, as well as measurements at R-band, in the lab
are also summarized in Table 1. The absolute flux normalization at 7.8m
is given in Table 2.
We then made measurements of the calibrated light source through the same
filter set from the roof of Swain Hall and the roof of the Indiana University
Law Library, which was surveyed by the Indiana Geological survey to be 124.1m
distant. In principle, the flux from the lamp should illuminate the
telescope like the light from the star. At 124.1m, the 4mm lamp filament
has a angular size of 6.6". Although the signal strength was sufficient
to move the source farther back, thre was no way to do this without falling
off the roof. We do not believe this to have a significant effect in
our measurements. These measurements were made over a period of 6 weeks
and were quite repeatable.
The results of the 1/R**2 mesurements between the Swain Hall roof and the
Law Library roof are given in Table 1. In all bandpasses there is a
clear and significant deficit relative to the expected 1/R**2 law over a 124.1m
baseline. At 1050nm in particular, there is no obvious reason for this
discrepancy. But the deficit clearly grows as the atmospheric opacity
grows, as it should. The sightline is bordered on the north by woods
and on the south by a well-travelled street. It passes directly over
a fraternity house, although the house is currently vacant during the summer
recess.
Table 1
The flux ratio, F/F0, measured in the lab at 7.8m and 8.5m, and on the Law
Library roof at 124.1m through filters centered at 1050nm, 1200nm, 1400nm,
and R-band normalized to the fiducial flux at 7.8m. The table gives
for each filter at 8.5m and 124.1m the measured ratio (meas), the expected
ratio (exp), and the fractional deviation (D)
|
F/Fo
|
|
|
7.8m
|
8.5m
|
124.1m
|
|
Filter
|
fid
|
meas.
|
exp.
|
D
|
meas.
|
exp.
|
D
|
|
1050nm
|
1
|
0.825
+ 0.009
|
0.842
|
-0.020
|
0.003266
+ 0.00006
|
0.00398
|
-0.179
|
|
1200nm
|
1
|
0.834
+ 0.016
|
″
|
-0.009
|
0.003181
+ 0.00015
|
″
|
-0.201
|
|
1400nm
|
1
|
0.823
+ 0.029
|
″
|
-0.023
|
0.001460
+ 0.00005
|
″
|
-0.633
|
|
R-Band
|
1
|
0.826
+ 0.010
|
″
|
-0.019
|
0.002346
|
″
|
-0.411
|
Table 2. The absolute flux measured
at 7.8m in the filter bandpasses
|
Filter
|
F (mW/m²/nm)
@ 7.8m
|
|
1050
|
0.03675
|
|
1200
|
0.03339
|
|
1400
|
0.02716
|
|
R-Band
|
0.01808
|
Fig.4(a) and 4(b) below show the experimental results described above
in graphical form. At the 'near' pathlengths, the results conform to
the 1/R**2 law. But at the 124.1m pathlength there is a significant
deviation from the 1/R**2 law, even at 1050nm where there is no water vapor
opacity in the bandpass.
Figure 4. (a) Graphical display of the results in Table 1. The
solid line reprents the 1/R**2 law, normalized to 1 at 7.8m. The 'near'
mesurements for all filters fall on the line. The 'far' mesurements
to the roof of the Law Library for all filters clearly fall below the line.
Figure 4. (b) Enlarged view of the deviation of the 'far'
measurements from the 1/R**2 law. Statistical errors in the measurements
are small compared with the size of the symbol. The deviations are statistically
significant and the deviations increase as the atmospheric absorption increases.
Conclusion: If this approach to the calibration of
fundamental SNAP photometric standards is to succeed, this experiment must
be repeated at better sites, and over a larger range of pathlengths, to prove
whether the 1/R**2 law holds. If environmental effects are responsible
for our observed 1/R**2 deficit at 1050 nm, futher measurements at better
sites should validate this piece of the calibration procedure.
I. Tests using the Moon
In addition to the 1/R**2 tests described above, we have been testing
whether observations of stars would be free of opacity in the narrow bandpass
filters in the near infrared. One problem we faced was the fact that
our photometer was not an integrating device. Basically we just read
a voltage. So we are unable to measure stars, even a star as bright
as Arcturus in the near infrared. For a few $100 we could have made
an integrator, but we chose a different and cheaper route. We looked
at the Moon as a function of zenith angle. This should tell us whether
our bandpasses are clear. So far our results are problematical. But
these results were obtained over only a 3 day period and our observations
are continuing. We have had uncooperative weather over the past 3 days
running. (Indiana, go figure)
In Figs.5-7, we show the results of measurements of the Moon through filters
at 1050nm, 1200nm, and 1400nm as a function of zenith angle. The procedure
is to align the Meade on a particular feature on the Moon and allow the telescope
to track (it has a Moon tracking program). We measured the Moon through
each filter twice for 1 minute with 50 ms samples, and averaged the result.
We then cycled to the next filter. The instantaneous zenith angle
was determined from Voyager III. The Meade's ability to track the Moon
is compromised somewhat, however, because the torque on the gears from the
photometer is clearly much greater than the motors have been designed for,
even though we wrapped the front end of the telescope tube with lead sheet.
The gears now obviously have backlash that they didn't have when we
began these measurements.
Except for transit on August 9, the results are consistent. The
1050nm bandpass appears clear, the 1200nm bandpass shows the effects of some
opacity, and the 1400nm bandpass is affected most by opacity. What happened
to double the signal on August 9 near transit? We do not understand
this now and only more measurements will enable us to decide whether the effect
is real or a systematic introduced by our less than robust setup.
Figure 5. Measurements of the Moon over 3 days in the 1050nm
filter as a function of sec(zentih angle)
Figure 6. Measurements of the Moon over 3 days in the 1200nm
filter as a function of the sec(zenith angle)
Figure 7. Measurements of the Moon over 3 days in the 1400nm
filter as a function of the sec(zenith angle)